DNA, RNA, and protein strongly absorb ultraviolet light in the 260 to 280 nm range. UV spectroscopy can be used as a quantitative technique to measure nucleic acid concentration and protein contamination. Nucleic acids strongly absorb at 260 nm and less strongly at 280 nm while proteins do the opposite. The general rules for determining the concentrations of nucleic acids at 260 nm are:

  1. 1 Optical Density (OD) unit of double-stranded DNA is 50 micrograms/ml.
  2. 1 OD unit of single-stranded DNA is 33 micrograms/ml.
  3. 1 OD unit of single-stranded RNA is 40 micrograms/ml.

Proteins absorb strongly at 280 nm where 1 OD unit is 1 mg/ml. When using UV spectroscopy for estimating DNA concentrations, it is very important to remove all protein and RNA from the DNA solution. Good estimations can only be made on clean preparations.

An estimate of the purity of a DNA preparation can be made by measuring the absorbance at both 260 nm and 280 nm. Pure solutions of nucleic acid will absorb approximately twice as much at 260 nm as at 280 nm. Experimentally, the ratio of 260 nm/280 nm of a pure DNA solution is between 1.8 to 2.0. As protein contamination increases, the ratio decreases. Additionally, the presence of contaminating oranic solvents, such as phenol, can affect estimations of concentration and purity.

Materials you need are:

UV Spectrophotometer
Quartz or UV compatible cuvettes
TE buffer
DNA template

Method:

  1. Fill the cuvette with water or TE buffer. Zero the spectrophotometer at 260 nm with this blank.
  2. DNA from plasmid and genomic preparations is typically at a concentration exceeding 1 micrograms/microliter. Consequently, DNA is usually diluted before measuring its absorbance. An unfortunate result of this measurement is that the DNA is expended as a result of the dilution. Be sure these is adequate DNA to waste. Start by diluting the DNA sample 1 microliter : 999 microliters of TE buffer (the dilution can be done directly in the cuvette). Mix the dilution thoroughly.
  3. Measure the optical density (OD). Multiply the resulting OD by 50 micrograms/ml. For a 1:1000 dilution, the mass of DNA is equal to micrograms/microliter.
  4. Similarly, the same sample can be measured at 280 nm. A ratio of the OD-260nm/OD-280nm is an indicator of DNA purity. A ratio of 1.8 or higher indicates minimal protein contamination.

Measuring DNA concentration is not difficult, is it?

Regards

advertisement

Incoming Search Terms : estimation of dna,quantitative estimation of dna,DNA estimation,estimation of DNA concentration,dna estimation using spectrophotometer,determine protein purity by spectrophotometry,od dna concentration,DNA purity ratio,Quantitative Estimation of DNA Concentrations,estimation of dna and rna,dna purity measurement,plasmid OD260/230,dna concentration spectrophotometer,nucleic acid estimation,Uv determination of DNA,estimation of dna by spectroscopy,dna quantitative estimation,calculation of DNA concentration in TE buffer,1 abs unit=50 microgram/ml DNA,1 abs= 50 microgram of PLASMID DNA,dna concentrations,OD 260/230 DNA,DNA od,dna purity absorbance,dna purity and concentration,RNA OD A260/A230 ratio for qPCR,Does TE buffer affect UV absorbance of DNA,single stranded dna od ratio,specrophotometric determination of plasmid concentration,dna quantitative in spectrophotometer

2 Responses leave one →
  1. Eboireau permalink
    June 21, 2010

    You can miniaturize DNA/RNA assay by UV (and also using nucleic acids stains such as Hoechts dye) by performing the UV abs measure in IMAplates with a UV microplate reader: up 96 samples of 1-6µl can be measured ! See http://www.interchim.fr/ft/B/BA361n.pdf and application of DNA dosage on 1-6µl at http://www.interchim.fr/ft/B/BA361d.pdf

  2. Eboireau permalink
    June 21, 2010

    UV absorbance (of DNA/RNA samples) can be measured in volumes down 6µl and even 1µl using the IMAplates with a standard UV microplate reader.
    Samples can be loaded manually, up 96 samples in 5-10seconds! See http://www.interchim.fr/ft/B/BA361n.pdf and application of DNA dosage of DNA at http://www.interchim.fr/ft/B/BA361d.pdf

Leave a Reply

Note: You can use basic XHTML in your comments. Your email address will never be published.

Subscribe to this comment feed via RSS